Introduction
In
early 1900s, Gas chromatography (GC) was discovered by Mikhail
Semenovich Tsvett as a separation technique to separate compounds. In
organic chemistry, liquid-solid column chromatography is often used to
separate organic compounds in solution. Among the various types of gas
chromatography, gas-liquid chromatography is the method most commonly
used to separate organic compounds. The combination of gas
chromatography and mass spectrometry is an invaluable tool in the
identification of molecules. A typical gas chromatograph consists of an
injection port, a column, carrier gas flow control equipment, ovens and
heaters for maintaining temperatures of the injection port and the
column, an integrator chart recorder and a detector.
To
separate the compounds in gas-liquid chromatography, a solution sample
that contains organic compounds of interest is injected into the sample
port where it will be vaporized. The vaporized samples that are injected
are then carried by an inert gas, which is often used by helium or
nitrogen. This inert gas goes through a glass column packed with silica
that is coated with a liquid. Materials that are less soluble in the
liquid will increase the result faster than the material with greater
solubility.The purpose of this module is to provide a better
understanding on its separation and measurement techniques and its
application.
In GLC, the liquid
stationary phase is adsorbed onto a solid inert packing or immobilized
on the capillary tubing walls. The column is considered packed if the
glass or metal column tubing is packed with small spherical inert
supports. The liquid phase adsorbs onto the surface of these beads in a
thin layer. In a capillary column, the tubing walls are coated with
the stationary phase or an adsorbant layer, which is capable of
supporting the liquid phase. However, the method of GSC, has limited
application in the laboratory and is rarely used due to severe peak
tailing and the semi-permanent retention of polar compounds within the
column. Therefore, the method of gas-liquid chromatography is simply
shortened to gas chromatography and will be referred to as such here.The
purpose of this module is to provide a better understanding on its
separation and measurement techniques and its application.
Instrumentation
Sample Injection
A sample port is necessary for
introducing the sample at the head of the column. Modern injection
techniques often employ the use of heated sample ports through which the
sample can be injected and vaporized in a near simultaneous fashion. A
calibrated microsyringe is used to deliver a sample volume in the range
of a few microliters through a rubber septum and into the vaporization
chamber. Most separations require only a small fraction of the initial
sample volume and a sample splitter is used to direct excess sample to
waste. Commercial gas chromatographs often allow for both split and
splitless injections when alternating between packed columns and
capillary columns. The vaporization chamber is typically heated 50 °C
above the lowest boiling point of the sample and subsequently mixed with
the carrier gas to transport the sample into the column.
Figure 1. A cross-sectional view of a microflash vaporizer direct injector.
Carrier Gas
The carrier gas plays an important
role, and varies in the GC used. Carrier gas must be dry, free of oxygen
and chemically inert mobile-phase employed in gas chromatography.
Helium is most commonly used because it is safer than, but comprable to
hydrogen in efficiency, has a larger range of flow rates and is
compatable with many detectors. Nitrogen, argon, and hydrogen are also
used depending upon the desired performance and the detector being
used. Both hydrogen and helium, which are commonly used on most
traditional detectors such as Flame Ionization(FID), thermal
conductivity (TCD) and Electron capture (ECD), provide a shorter
analysis time and lower elution temperatures of the sample due to higher
flow rates and low molecular weight. For instance, hydrogen or helium
as the carrier gas gives the highest sensitivity with TCD because the
difference in thermal conductivity between the organic vapor and
hydrogen/helium is greater than other carrier gas. Other detectors such
as mass spectroscopy, uses nitrogen or argon which has a much better
advantage than hydrogen or helium due to their higher molecular weights,
in which improve vacuum pump efficiency.
All carrier gasses are available in
pressurized tanks and pressure regulators, gauges and flow meters are
used to meticulously control the flow rate of the gas. Most gas supplies
used should fall between 99.995% - 99.9995% purity range and contain a
low levels (< 0.5 ppm) of oxygen and total hydrocarbons in the tank.
The carrier gas system contains a molecular sieve to remove water and
other impurities. Traps are another option to keep the system pure and
optimum sensitive and removal traces of water and other contaminants. A
two stage pressure regulation is required to use to minimize the
pressure surges and to monitor the flow rate of the gas. To monitor the
flow rate of the gas a flow or pressure regulator was also require onto
both tank and chromatograph gas inlet. This applies different gas type
will use different type of regulator.The carrier gas is preheated and
filtered with a molecular sieve to remove impurities and water prior to
being introduced to the vaporization chamber. A carrier gas is typically
required in GC system to flow through the injector and push the gaseous
components of the sample onto the GC column, which leads to the
detector ( see more detail in detector section).
Figure 2. Gas
Recommendations for Capillary
Columns Figure 3. Gas
Recommendations for Packed Columns
Column Oven
The thermostatted oven serves to control
the temperature of the column within a few tenths of a degree to
conduct precise work. The oven can be operated in two
manners: isothermal programming or temperature programming. In
isothermal programming, the temperature of the column is held constant
throughout the entire separation. The optimum column temperature for
isothermal operation is about the middle point of the boiling range of
the sample. However, isothermal programming works best only if the
boiling point range of the sample is narrow. If a low isothermal column
temperature is used with a wide boiling point range, the low boiling
fractions are well resolved but the high boiling fractions are slow to
elute with extensive band broadening. If the temperature is increased
closer to the boiling points of the higher boiling components, the
higher boiling components elute as sharp peaks but the lower boiling
components elute so quickly there is no separation.
In the temperature programming method,
the column temperature is either increased continuously or in steps as
the separation progresses. This method is well suited to separating a
mixture with a broad boiling point range. The analysis begins at a low
temperature to resolve the low boiling components and increases during
the separation to resolve the less volatile, high boiling components of
the sample. Rates of 5-7 °C/minute are typical for temperature
programming separations.
Figure 3. The effect of column temperature on the shape of the peaks.
Open Tubular Columns and Packed Columns
Open tubular columns, which are also
known as capillary columns, come in two basic forms. The first is a
wall-coated open tubular (WCOT) column and the second type is a
support-coated open tubular (SCOT) column. WCOT columns are capillary
tubes that have a thin later of the stationary phase coated along the
column walls. In SCOT columns, the column walls are first coated with a
thin layer (about 30 micrometers thick) of adsorbant solid, such as
diatomaceous earth, a material which consists of single-celled,
sea-plant skeletons. The adsorbant solid is then treated with the
liquid stationary phase. While SCOT columns are capable of holding a
greater volume of stationary phase than a WCOT column due to its greater
sample capacity, WCOT columns still have greater column efficiencies.
Most modern WCOT columns are made of
glass, but T316 stainless steel, aluminum, copper and plastics have also
been used. Each material has its own relative merits depending upon
the application. Glass WCOT columns have the distinct advantage of
chemical etching, which is usually achieved by gaseous or concentrated
hydrochloric acid treatment. The etching process gives the glass a
rough surface and allows the bonded stationary phase to adhere more
tightly to the column surface.
One of the most popular types of
capillary columns is a special WCOT column called the fused-silica
wall-coated (FSWC) open tubular column. The walls of the fused-silica
columns are drawn from purified silica containing minimal metal oxides.
These columns are much thinner than glass columns, with diameters as
small as 0.1 mm and lengths as long as 100 m. To protect the column, a
polyimide coating is applied to the outside of the tubing and bent into
coils to fit inside the thermostatted oven of the gas chromatography
unit. The FSWC columns are commercially available and currently
replacing older columns due to increased chemical inertness, greater
column efficiency and smaller sampling size requirements. It is
possible to achieve up to 400,000 theoretical plates with a 100 m WCOT
column, yet the world record for the largest number of theoretical
plates is over 2 million plates for 1.3 km section of column.
Packed columns are made of a glass or a
metal tubing which is densely packed with a solid support like
diatomaceous earth. Due to the difficulty of packing the tubing
uniformly, these types of columns have a larger diameter than open
tubular columns and have a limited range of length. As a result, packed
columns can only achieve about 50% of the efficiency of a comparable
WCOT column. Furthermore, the diatomaceous earth packing is deactivated
over time due to the semi-permanent adsorption of impurities within the
column. In contrast, FSWC open tubular columns are manufactured to be
virtually free of these adsorption problems.
Figure 4. Properties of gas chromatography columns.
Figure 5. Computer Generated Image of a FSWC column (specialized for measuring BAC levels)
Figure 6. Computer Generated Image of a FSWC column (specialized to withstand extreme heat)
Different types of columns can be
applied for different fields. Depending on the type of sample, some GC
columns are better than the others. For example, the FSWC column shown
in Figure 5 is designed specially for blood alcohol analysis. It
produces fast run times with baseline resolution of key components in
under 3 minutes. Moreover, it displays enhanced resolutions of ethanol
and acetone peaks, which helps with determining the BAC levels. This
particular column is known as Zebron-BAC and it made with polyimide
coating on the outside and the inner layer is made of fused silica and
the inner diameter ranges from .18 mm to .25 mm. There are also many
other Zebron brand columns designed for other purposes.
Another example of a Zebron GC column is
known as the Zebron-inferno. Its outer layer is coated with a special
type of polyimide that is designed to withstand high temperatures. As
shown in figure 6, it contains an extra layer inside. It can withstand
up to 430 °C to be exact and it is designed to provide true boiling
point separation of hydrocarbons distillation methods. Moreover, it is
also used for acidic and basic samples.
Detection Systems
The detector is the device located at
the end of the column which provides a quantitative measurement of the
components of the mixture as they elute in combination with the carrier
gas. In theory, any property of the gaseous mixture that is different
from the carrier gas can be used as a detection method. These detection
properties fall into two categories: bulk properties and specific
properties. Bulk properties, which are also known as general
properties, are properties that both the carrier gas and analyte possess
but to different degrees. Specific properties, such as detectors that
measure nitrogen-phosphorous content, have limited applications
but compensate for this by their increased sensitivity.
Each detector has two main parts that
when used together they serve as transducers to convert the detected
property changes into an electrical signal that is recorded as a
chromatogram. The first part of the detector is the sensor which is
placed as close the the column exit as possible in order to optimize
detection. The second is the electronic equipment used to digitize the
analog signal so that a computer may analyze the acquired chromatogram.
The sooner the analog signal is converted into a digital signal, the
greater the signal-to-noise ratio becomes, as analog signal are easily
susceptible to many types of interferences.
An ideal GC detector is distinguished by
several characteristics. The first requirement is adequate sensitivity
to provide a high resolution signal for all components in the mixture.
This is clearly an idealized statement as such a sample would approach
zero volume and the detector would need infinite sensitivity to detect
it. In modern instruments, the sensitivities of the detectors are in
the range of 10-8 to 10-15 g of solute per
second. Furthermore, the quantity of sample must be reproducible and
many columns will distort peaks if enough sample is not injected. An
ideal column will also be chemically inert and and should not alter the
sample in any way. Optimized columns will be able to withstand
temperatures in the range of -200 °C to at least 400 °C. In addition,
such a column would have a short linear response time that is
independent of flow rate and extends for several orders of magnitude.
Moreover, the detector should be reliable, predictable and easy to
operate.
Understandably, it is not possible for a
detector meet all of these requirements. The next subsections will
discuss some of the more common types of gas chromatography detectors
and the relative advantages and/or disadvantages of each.
Type of Detector
|
Applicable Samples
|
Detection Limit
|
Mass Spectrometer (MS)
|
Tunable for any sample
|
.25 to 100 pg
|
Flame Ionization (FID)
|
Hydrocarbons
|
1 pg/s
|
Thermal Conductivity (TCD)
|
Universal
|
500 pg/ml
|
Electron-Capture (ECD)
|
Halogenated hydrocarbons
|
5 fg/s
|
Atomic Emission (AED)
|
Element-selective
|
1 pg
|
Chemiluminescence (CS)
|
Oxidizing reagent
|
Dark current of PMT
|
Photoionization (PID)
|
Vapor and gaseous Compounds
|
.002 to .02 µg/L
|
Table 7: Typical gas chromatography detectors and their detection limits.
Mass Spectrometry Detectors
Mass Spectrometer (MS) detectors are most
powerful of all gas chromatography detectors. In a GC/MS system, the
mass spectrometer scans the masses continuously throughout the
separation. When the sample exits the chromatography column, it
is passed through a transfer line into the inlet of the mass
spectrometer . The sample is then ionized and fragmented, typically by
an electron-impact ion source. During this process, the sample is
bombarded by energetic electrons which ionize the molecule by causing
them to lose an electron due to electrostatic repulsion. Further
bombardment causes the ions to fragment. The ions are then passed into a
mass analyzer where the ions are sorted according to their m/z value,
or mass-to-charge ratio. Most ions are only singly charged.
The Chromatogram will point out the
retention times and the mass spectrometer will use the peaks to
determine what kind of molecules are exist in the mixture. The figure
below represents a typical mass spectrum of water with the absorption
peaks at the appropriate m/z ratios.
Figure 8. Mass Spectrum of Water
Instrumentation
One of the most common types of mass
analyzer in GC/MS is the quadrupole ion-trap analyzer, which allows
gaseous anions or cations to be held for long periods of time by
electric and magnetic fields. A simple quadrupole ion-trap consists of a
hollow ring electrode with two grounded end-cap electrodes as seen in
figure #. Ions are allowed into the cavity through a grid in the upper
end cap. A variable radio-frequency is applied to the ring electrode and
ions with an appropriate m/z value orbit around the cavity. As the
radio-frequency is increased linearly, ions of a stable m/z value are
ejected by mass-selective ejection in order of mass. Ions that are too
heavy or too light are destabilized and their charge is neutralized upon
collision with the ring electrode wall. Emitted ions then strike an
electron multiplier which converts the detected ions into an electrical
signal. This electrical signal is then picked up by the computer through
various programs. As an end result, a chromatogram is produced
representing the m/z ratio versus the abundance of the sample.
GC/MS units are advantageous because
they allow for the immediate determination of the mass of the analyte
and can be used to identify the components of incomplete separations.
They are rugged, easy to use and can analyze the sample almost as
quickly as it is eluted. The disadvantages of mass spectrometry
detectors are the tendency for samples to thermally degrade before
detection and the end result of obliterating all the sample by
fragmentation.
Figure 9. Schematic of the GC/MS system.
Figure 10. Arrangement of the poles in Quadrupole and Ion Trap Mass spectrometers
Flame Ionization Detectors
Flame ionization detectors (FID) are the
most generally applicable and most widely used detectors. In a FID, the
sample is directed at an air-hydrogen flame after exiting the column.
At the high temperature of the air-hydrogen flame, the sample undergoes
pyrolysis, or chemical decomposition through intense heating. Pyrolized
hydrocarbons release ions and electrons that carry current. A
high-impedance picoammeter measures this current to monitor the sample's
elution.
It is advantageous to used FID because
the detector is unaffected by flow rate, noncombustible gases and
water. These properties allow FID high sensitivity and low noise. The
unit is both reliable and relatively easy to use. However, this
technique does require flammable gas and also destroys the sample.
Figure 11. Schematic of a typical flame ionization detector.
Thermal Conductivity Detectors
Thermal conductivity detectors (TCD) were
one the earliest detectors developed for use with gas chromatography.
The TCD works by measuring the change in carrier gas thermal
conductivity caused by the presence of the sample, which has a different
thermal conductivity from that of the carrier gas. Their design is
relatively simple, and consists of an electrically heated source that is
maintained at constant power. The temperature of the source depends
upon the thermal conductivities of the surrounding gases. The source is
usually a thin wire made of platinum, gold or . The resistance within
the wire depends upon temperature, which is dependent upon the thermal
conductivity of the gas.
TCDs usually employ two detectors, one of
which is used as the reference for the carrier gas and the other which
monitors the thermal conductivity of the carrier gas and sample
mixture. Carrier gases such as helium and hydrogen has very high
thermal conductivities so the addition of even a small amount of sample
is readily detected.
The advantages of TCDs are the ease and
simplicity of use, the devices' broad application to inorganic and
organic compounds, and the ability of the analyte to be collected after
separation and detection. The greatest drawback of the TCD is the low
sensitivity of the instrument in relation to other detection methods, in
addition to flow rate and concentration dependency.
Figure 12. Schematic of thermal conductivity detection cell.
Figure 13. Standard Chromatogram of a Mixture of Gases
Chromatogram
Figure 13 represents a standard
chromatogram produced by a TCD detector. In a standard chromatogram
regardless of the type detector, the x-axis is the time and the y-axis
is the abundance or the absorbance. From these chromatograms, retention
times and the peak heights are determined and used to further
investigate the chemical properties or the abundance of the samples.
Electron-capture Detectors
Electron-capture detectors (ECD) are
highly selective detectors commonly used for detecting environmental
samples as the device selectively detects organic compounds
with moieties such as halogens, peroxides, quinones and nitro groups and
gives little to no response for all other compounds. Therefore, this
method is best suited in applications where traces quantities of
chemicals such as pesticides are to be detected and other
chromatographic methods are unfeasible.
The simplest form of ECD involves gaseous
electrons from a radioactive ? emitter in an electric field. As the
analyte leaves the GC column, it is passed over this ? emitter, which
typically consists of nickle-63 or tritium. The electrons from the ?
emitter ionize the nitrogen carrier gas and cause it to release a burst
of electrons. In the absence of organic compounds, a constant standing
current is maintained between two electrodes. With the addition of
organic compounds with electronegative functional groups, the current
decreases significantly as the functional groups capture the
electrons.
The advantages of ECDs are the high
selectivity and sensitivity towards certain organic species with
electronegative functional groups. However, the detector has a limited
signal range and is potentially dangerous owing to its radioactivity.
In addition, the signal-to-noise ratio is limited by radioactive decay
and the presence of O2 within the detector.
Figure 14. Schematic of an electron-capture detector.
Atomic Emission Detectors
Atomic emission detectors (AED), one of
the newest addition to the gas chromatographer's arsenal, are
element-selective detectors that utilize plasma, which is a partially
ionized gas, to atomize all of the elements of a sample and excite their
characteristic atomic emission spectra. AED is an extremely powerful
alternative that has a wider applicability due to its based on the
detection of atomic emissions.There are three ways of generating plasma:
microwave-induced plasma (MIP), inductively coupled plasma (ICP) or
direct current plasma (DCP). MIP is the most commonly employed form and
is used with a positionable diode array to simultaneously monitor the
atomic emission spectra of several elements.
InstrumentationThe components of the Atomic emission detectors include 1) an interface for the incoming capillary GC column to induce plasma chamber,2) a microwave chamber, 3) a cooling system, 4) a diffration grating that associated optics, and 5) a position adjustable photodiode array interfaced to a computer.
Figure 15. Schematic of atomic emission detector.
GC Chemiluminescence Detectors
Chemiluminescence spectroscopy (CS) is a
process in which both qualitative and quantitative properties can be be
determined using the optical emission from excited chemical species. It
is very similar to AES, but the difference is that it utilizes the
light emitted from the energized molecules rather than just excited
molecules. Moreover, chemiluminescence can occur in either the solution
or gas phase whereas AES is designed for gaseous phases. The light
source for chemiluminescence comes from the reactions of the chemicals
such that it produces light energy as a product. This light band is used
instead of a separate source of light such as a light beam.
Like other methods, CS also has its limitations and the major
limitation to the detection limits of CS concerns with the use of a
photomultiplier tube (PMT). A PMT requires a dark current in it to
detect the light emitted from the analyte.
Figure 16. Schematic of a GC Chemiluminescence Detector
Photoionization Detectors
Another different kind of detector for
GC is the photoionization detector which utilizes the properties of
chemiluminescence spectroscopy. Photoionization detector (PID) is a
portable vapor and gas detector that has selective determination of
aromatic hydrocarbons, organo-heteroatom, inorganice species and other
organic compounds. PID comprise of an ultrviolet lamp to emit photons
that are absorbed by the compounds in an ionization chamber exiting from
a GC column. Small fraction of the analyte molecules are actually
ionized, nondestructive, allowing confirmation analytical results
through other detectors. In addition, PIDs are available in portable
hand-held models and in a number of lamp configurations. Results are
almost immediate. PID is used commonly to detect VOCs in soil, sediment,
air and water, which is often used to detect contaminants in ambient
air and soil. The disavantage of PID is unable to detect certain
hydrocarbon that has low molecular weight, such as methane and ethane.
Instrumentation
Figure 17. Schematic of a photoionization detector
Limitations
- Not suitable for detecting semi-volatile compounds
- Only indicates if volatile organic compounds are presents.
- High concentration so methane are required for higher performance.
- Frequent calibration are required.
- Units of parts per million range
- Enviromental distraction, especially water vapor.
- Strong electrical fieldsRapid variation in temperature at the detector and naturally occurring compounds may affect instrumental signal.
Applications
Gas chromatography is a physical
separation method in where volatile mixtures are separated. It can be
used in many different fields such as pharmaceuticals, cosmetics and
even environmental toxins. Since the samples have to be volatile, human
breathe, blood, saliva and other secretions containing large amounts of
organic volatiles can be easily analyzed using GC. Knowing the amount of
which compound is in a given sample gives a huge advantage in studying
the effects of human health and of the environment as well.
Air samples can be analyzed using GC.
Most of the time, air quality control units use GC coupled with FID in
order to determine the components of a given air sample. Although other
detectors are useful as well, FID is the most appropriate because of its
sensitivity and resolution and also because it can detect very small
molecules as well.
GC/MS is also another useful method
which can determine the components of a given mixture using the
retention times and the abundance of the samples. This method be applied
to many pharmaceutical applications such as identifying the amount of
chemicals in drugs. Moreover, cosmetic manufacturers also use this
method to effectively measure how much of each chemical is used for
their products.
Equations
“Height equivalent to a theoretical
plate” (HETP) use to calculate the flow rate by usingthe total number
of theoretical plates (N) and column length (L). Some application, HETP
concepts is used in industrial practice to convert number of theoretical
plates to packing height. HETP can be calculate with the Van Deemter
equation, which is given by
Where A and B and C are constants and v is the linear velocity (carrier flow rate).
- A is the "Eddy-Diffusion" term and causes the broadering of the solute band.
- B is the "Longitudinal diffusion" term whereby the concentration of the analyte, in which diffuses out from the center to the edges.This causes the broadering of the analyte band.
- C is the "Resistance to Mass Transfer " term and causes the band of the analyte broader.
L is the length of the column, where N
is the number of theoretical plates, tR is the retention time, and ω is
the width of the elution peak at its base.
In which, the more plates give a better resolution and more efficiency. Resolution can be determined by
R= 2[ (tR)B – (tR)A] / ( WA +WB) (4)
A relationship between the plates and resolution is giving by,
No comments:
Post a Comment